Coulometric Sensor Method Oxygen Permeability Analyzer
GBPI Barrier Testing Equipment
Coulometric Sensor Method Oxygen Permeability Analyzer Y310L
Looking for technical specs, quotation or application guidance?
Get complimentary support from our professional testing instrument specialists.
Click for Free Consultation
Introduction
This product is designed and manufactured based on the coulometric oxygen sensor testing principle and in accordance with standards such as GB/T 19789 and ASTM D3985. It utilizes a high‑precision imported oxygen sensor to ensure accurate test results. It is suitable for testing the oxygen transmission properties of films, sheets, paper, packaging, and related materials across industries such as food, pharmaceuticals, medical devices, daily chemicals, and photovoltaic electronics, providing efficient, wide‑range oxygen transmission rate testing for materials with high, medium, and low oxygen barrier properties.
Test Principle
The Y310L Oxygen Transmission Rate Tester operates based on the coulometric principle. A prepared specimen is secured between the two halves of the test chamber; high‑purity oxygen flows through the upper chamber, while high‑purity nitrogen (carrier gas) flows through the lower chamber. Oxygen molecules permeate the film into the nitrogen stream and are carried to the sensor. The oxygen transmission rate is then calculated by analyzing the oxygen concentration measured by the sensor.
Schematic Diagram for Coulometric Sensor Method
Applicable Standards
ISO 15105‑2, GB/T 19789, GB/T 1038.2, ASTM D3985, ASTM F2622, ASTM F1927, ASTM F1307, JIS K 7126‑2, DIN 53380‑3
Chinese Pharmacopoeia 2025 Edition "4007 Determination of Gas Transmission Rate of Pharmaceutical Packaging Materials", YBB 00082003
Specification
Item
Technical Parameter
Test Range
0.05~1000 cm³/(m²·24h·0.1MPa)
Resolution
0.0001 cm³/(m²·24h·0.1MPa)
Temperature Control Range
15~45℃
Temperature Control Precision
±0.2℃
Humidity Control Range
0%RH, 5~90%RH, 100%RH
Humidity Control Precision
±2%RH
Permeation Area
50.24 cm² (Custom‑made adapters, as small as 0.785 cm²)
Specimen Dimensions
Φ100 mm
Specimen Thickness
≤3 mm
Number of Tests
3 pcs
Carrier Gas Pressure
≥0.1 MPa
Carrier Gas Flow Rate
5~100 mL/min
External Dimensions
700 mm × 655 mm × 535 mm
Weight
60 kg
Power
750 W
Power Supply
AC 220 V, 50 Hz
Features
High‑Precision Oxygen Sensor
Features a high‑sensitivity oxygen sensor with exceptional stability and an ultra‑low failure rate; resolution reaches 0.0001 cm³/(m²·24h·0.1MPa).
Accurate and Reliable Test Data
Precise control of test temperature and humidity. Equipped with high‑precision temperature and humidity sensors, the system ensures stable conditions within the test chambers, with temperature control accuracy of 0.2°C and humidity precision of ±2% RH. Automated testing mode monitors temperature, humidity, and gas flow values and curves in real‑time, ensuring high test repeatability.
Independent Three‑Chamber Design for High Efficiency
Equipped with three chambers—each with a standard testing area of 50.24 cm²—the instrument meets requirements for testing three parallel samples simultaneously, ensuring high efficiency. Supports simultaneous testing of three identical or different samples, with independent data for each. Wide measurement range (0.05–1000 cm³/(m²·24h·0.1MPa)) accommodates materials with high, medium, and low barrier properties; optional adapters allow for oxygen transmission testing of containers such as bottles, bags, and bowls.
Intelligent Operating System with Global Certification
Proprietary intelligent operating system featuring a modular, graphical interface; allows for flexible parameter settings and intuitive, convenient operation. Designed in compliance with GMP Annex "Computerized Systems," featuring audit trails and multi‑level user permissions to meet pharmaceutical industry requirements for data traceability. Customizable test reports; supports multiple data output formats, electronic signatures, and online report submission for review.
Professional Calibration Services for Accurate and Reliable Data
We hold the "National Reference Material Grading Certificate" for gas transmission rates, approved and issued by the General Administration of Quality Supervision, Inspection and Quarantine of the P.R.C. (Reference Material Nos. GBW(E) 130497 / GBW(E) 130498). Instruments are calibrated and verified using these national reference materials, ensuring the accuracy, universal applicability, and authoritative nature of the test data.
Application
Application Field
Description
Film
Oxygen transmission rate test of various plastic films (PP/PET/PE/PVC/BOPP/CPP, etc.), plastic composite films, paper‑plastic composite films, metal composite films, co‑extrusion films, aluminized films, degradable packaging films (PLA/PBAT/PBS, etc.) and other film‑like materials.
Sheet
Oxygen transmission rate test of solid pharmaceutical hard sheets (PP/PVC/PTP, etc.), metal composite sheets, rubber sheets and other flakes.
Paper, Cardboard and Its Composites
Oxygen transmission rate test of coated paper, silicone paper, cigarette bag aluminized paper, paper aluminum‑plastic composite sheet and other paper and cardboard.
About GBPI
GBPI (Guangzhou Biaoji Packaging Equipment Co., Ltd.) was founded in 2002 in Guangzhou Economic and Technological Development Zone. It is a comprehensive high‑tech enterprise integrating R&D, manufacturing and sales of packaging & barrier testing instruments. We deliver complete testing solutions for packaging, food, pharmaceutical, medical device, new‑material and laboratory industries. GBPI owns a CNAS‑authorized testing laboratory and reference‑material development center, supplying instrument hardware, third‑party sample testing and calibration services for global clients.
Interested in the Y310L Oxygen Permeability Analyzer for your laboratory?
Reach out to GBPI for datasheet, quotation and professional technical support.